The physical size and power requirements are relatively nominal and are shown  below:

 Moisture Analyzer

Physical Properties and Utilities

Physical Information

Dimensions with gas cell
  • Length -  39 inches
  • Width -    23 inches
  • Height -  10 inches
  • Weight - 200 pounds
Gas Cell
  • Pathlength 6.1 meters
  • Volume 0.5 liters
  • Materials 304 stainless steel on all wetted materials by sample gas
  • Windows KBr (potassium bromide) 25x 4 mm
  • Conflat seals between metal-metal flanges
  • Viton O-rings between window and metal cell
  • Inlet/outlet 1/4 inch VCR male fittings, stainless steel, welded to flange
Power Requirements
110 volt, 6.5 amps serge on start up
Details of power requirements
FTIR power supply
  • 35 watts steady state  power drain
  • 110 volts 1.3 amperes maximum serge at start up
Gas cell heater blanket
110 volts, 4.6 amperes Maximum (high temperature cut-off)
 Controller electronics 110 volts, 0.6 amps
Maximum output from controller possible - 110 volts, 10 amps
TE-Cooled MCT Detector
  • 110 volts, 0.5 amps maximum serge on start up supply provides
  • 9 volts, 3 amps, 27 watts steady drain

 

F2 Impurity Analyzer

Physical Properties and Utilities

Physical Information

Dimensions with gas cell
  • Length -39 inches
  • Width -  23 inches
  • Height - 10 inches
  • Weight -200 pounds
 Gas Cell
  • Path length 1.9 meters 6.1 meters set in steps of 0.5 meters
  • Volume 0.5 liters
  • Materials 304 stainless steel on all wetted materials by sample gas
  • Windows CaF2  25x 4 mm
  • Conflat seals between metal-metal flanges
  •  Teflon O-rings between window and metal cell
  • Inlet/outlet 1/4 inch VCR male fittings, stainless steel, welded to flange
Power Requirements
110 volt, 6.5 amps serge on start up
Details of power requirements
FTIR power supply
  • 35 watts steady state power drain
  • 110 volts 1.3 amperes maximum serge at start up
Gas cell heater blanket
  • 110 volts, 4.6 amperes Maximum (high temperature cut-off)
Controller electronics 110 volts, 0.6 amps
  • Maximum output from controller possible  - 110 volts, 10 amps
 TE Cooled MCT Detector
  • 110 volts, 0.5 amps maximum serge on start up supply provides
  • 9 volts, 3 amps, 27 watts steady drain

 

[Home] [New - FTIR Gas Analyzer] [Google Site Search] [FCM (FTIR Collection)] [Active Gas Analyzer] [Design Features] [Gas Cell Features] [Enclosure Features] [FTIR Advantages] [Software Advantages] [Impurities in Fluorine] [Utilities] [In-situ Analyzer] [Process Analyzer] [Fiber Optics Systems] [Short Path Gas Cell] [Long Path Gas Cell] [Optical Elements] [O-Rings and Seals] [Gas Quant Libraries] [Repair Used Equipment] [About] [Contact] [Privacy Statement] [Training and Consulting]